Interleaved B0-Mapping During Dynamic Creatine-CEST for Correction of Temporarily Fluctuating B0 Inhomogeneities During Plantar Flexion Exercise at 7T

Publikation: KonferenzbeitragPosterpeer-review

Abstract

Chemical Exchange Saturation Transfer is prone to inhomogeneities of the static magnetic field (B0). Hence, accurate frequency correction is mandatory for reliable quantification. Currently established B0 correction approaches assume B0 inhomogeneities to be static during CEST experiments, but this is questionable in the presence of subject motion and scanner instabilities. Thus, we propose three different dynamic B0 correction methods for CEST that can compensate for B0 instability for each Z-spectral point separately and compare them to three established static B0 correction approaches that apply the same frequency shift to all Z-spectral points in phantom and in vivo experiments.
OriginalspracheEnglisch
Seiten#3758
PublikationsstatusVeröffentlicht - 2017
Extern publiziertJa
VeranstaltungISMRM 25th Annual Meeting & Exhibition - Hawai'i Convention Center, Honolulu, USA/Vereinigte Staaten
Dauer: 22 Apr. 201727 Apr. 2017