Abstract
An electrothermal atomic absorption method (ETAAS) for the direct determination of trace elements (Cd, Cr, Cu, Mn, Se) both in blood fractions (erythrocytes, plasma and lymphocytes) and whole blood was developed. Zeeman background correction and graphite tubes with L'vov platforms were used. Samples were diluted with HNO3/Triton X-100 and pipetted directly into the graphite tube. Ashing, pretreatment and atomization steps were optimized carefully for the different fractions and elements applying different matrix modifiers for each element. For the lymphocyte fraction a multi-fold injection technique was applied. Low detection limits of the ETAAS method (Cd 0.13 μg/L, Cr 0.11 μg/L, Cu 0.52 μg/L, Mn 0.13 μg/L, Se 0.7 μg/L of whole blood) combined with small quantities of sample necessary for analysis allow determination of trace elements in this matrix. Verification of possible differences in the trace element status of humans was performed with statistical significance (P < 0.05). In addition, a contribution to the determination of normal values of essential elements was achieved. The method was applied for determination of trace elements in human blood and blood fractions of two groups (n = 50) different in health status. © Springer-Verlag 2000.
Originalsprache | Englisch |
---|---|
Seiten (von - bis) | 627-632 |
Seitenumfang | 6 |
Fachzeitschrift | Fresenius' Journal of Analytical Chemistry |
Jahrgang | 368 |
Ausgabenummer | 6 |
DOIs | |
Publikationsstatus | Veröffentlicht - 2000 |